Models PFP7 & PFP7/C
The Jenway Models PFP7 and PFP7/C (Clincal Version), are low temperature, single channel Flame Photometers, designed for the routine determinations of Sodium (Na) and Potassium (K). Additional filters are available for the measurement of Lithium (Li), Barium (Ba) and Calcium (Ca).
Both instruments are fitted with automatic flame failure detection for user safety, making it ideal for use in clinical, industrial and educational applications.
The built-in linearizer enables readings of both Na and K, at the normal clinical serum concentrations, to be diluted 200:1 or 100:1 prior to presentation to the flame photometer. Specifications | Model | PFP7 | PFP7/C (Clinical Version) | | Ranges | - | 120-160 mMol/L Na (linearized) | | Limits of Detection | Na ≤ 0.2ppm K ≤ 0.2ppm Li ≤ 0.25ppm Ca ≤ 15ppm Ba ≤ 30ppm | - - Li ≤ 0.25ppm Ca ≤ 15ppm Ba ≤ 30ppm | | Reproducibility | ≤ 1% CV for 20 consecutive samples (10ppm Na set to read 50) | | Linearity | < 2% (3ppm Na & K and 5ppm Li set to read 100) | | Specificity | Interference from Na. K and Li equal in conc to test element will be >0.5% | | Output | Nominal 1.00V per 100 reading | | Services Required | | | Electricity | 90 - 125 Vac or 190 - 250 Vac @ 50/60 Hz | | Air | Oil free dry air. 6 Litres/minute supplied at 1Bar (14 psig) | | Fuel Gas | Propane. Butane, Natural Gas or Liquid Petroleum Gas | | Dimensions (mm) wxdxh | 420 x 360 x 300 | | Weight (kg) | 8 | P/N 500094 Model PFP7 Flame Photometer supplied with Na and K filters, connecting hoses & clips, compressor plug, drain trap and manual
P/N 500200 Model PFP7/C Flame Photometer (clinical version) supplied with Na and K filters, connecting hoses & clips, compressor plug, drain trap and manual
Price: Call
|